Yahoo Poland Wyszukiwanie w Internecie

Search results

  1. The AED Plus® Administrator’s Guide is to be used by responsible medical authorities in conjunction with the ZOLL AED Plus Operator’s Guide (REF 9650-0300-01). The AED Plus is to be used by trained rescuers to provide emergency defibrillation.

  2. View and Download ZOLL AED PLUS service manual online. AED PLUS medical equipment pdf manual download.

  3. • Replace electrodes, if expired • After the ZOLL AED Plus is powered on and completes its self-test, the unit will issue an “Adult Pads” voice prompt or a “Pediatric Pads” voice prompt to indicate the type of electrodes connected to the unit.

  4. Battery Replacement. For AED Plus units running software version 5.32 or higher, replace batteries every 5 years or if unit prompts. For earlier software versions, replace batteries every 3 years and place a dated AED Plus battery replacement reminder label next to the On/Off button.

  5. Check for the green check ( ) in the status indicator window showing that the AED Plus is ready to use. Verify that electrodes pads are within their expiration date. Verify that batteries are within their expiration date. Typically 5 years from date of installation.

  6. • Do not use or stack the Fully Automatic AED Plus unit with other equipment. If the unit is used or stacked with other equipment, verify proper operation prior to use. • Keep the Fully Automatic AED Plus unit away from magnetic resonance imaging (MRI) equipment.

  7. 13 lip 2017 · A symbol of an hour glass indicates an expiration date. This date specifies when the AED accessory should be replaced. If an expiration date is shown, one should replace the accessory no later than that printed date. A symbol of a building indicates a manufactured date.

  1. Ludzie szukają również